21 Aug 2001 Accuracy of
Digivac Gauges
Series 100, 200, 800 and all gauges which read microns (millitorr)
0 to 9 microns +/- 1 micron
10 to 99 microns +/- 5% of reading +/- 1 micron
100 to 999 microns +/- 10% of reading +/- 5 micron
1000 to 1999 microns +/- 15% of reading +/- 50 micron
Wide range gauges (200 series) at readings above 1
torr
1.0 to 9.9 torr +/- 15% of reading +/- .1 torr
10 to 99 torr +/- 20% of reading +/- .2 torr
100 to 760 torr +/- 25% of reading +/- 10 torr Transducer
gauges (760 series)
1 to 199.9 torr +/- 1/4% of reading +/- .3 torr
1 to 760 torr +/- 1/4% of reading +/- 1 torr
Transducer
gauges (760 series) with decimal range and user zero trim
(neon and laser filling)
0 to 40.0 torr +/- 1/2 % of
reading +/- .1 torr
|