Any gauge that leaves our factory is tested under actual vacuum in our manifold and will be calibrated against a NIST traceable standard.
Series 100, 200, 800 and all gauges which read microns (millitorr)
0 to 9 microns +/- 1 microns
10 to 99 microns +/- 10% of reading
100 to 999 microns +/- 10% of reading
1000 to 1999 microns +/- 15% of reading
Wide range gauges (200 series) at readings above 1 Torr
1.0 to 9.9 torr +/- 15% of reading
10 to 160 torr +/- 50% of reading
160 to 760 torr +/- 25% of reading
Transducer gauges (760 series)
1 to 20 Torr +/- 0.5 Torr
21 to 760 Torr +/-2 Torr
Combination gauges (276/876 series)
0 to .001 Torr +/- 1 milliTorr
.010 to .099 Torr +/- 10% of reading
.100 to 2.5 Torr +/- 10% of reading
2.5 to 20 Torr +/- 0.5 Torr
21 to 760 Torr +/-2 Torr
Active Gauge Controller (201 series)
The accuracy of the vacuum measurement depends chiefly on the accuracy of the Active Gauge used. Below are the Accuracy in the primary range of interest. Refer to the Active Guage Data sheet for full range and accuracy over full range. We offer the ability to control 1-3 of the following Capacitance Manometer and active gauge:
Setra 730 Capacitance Manometer: 0.5% of Reading
MKS 722 Capacitance Manometer: 0.5% of Reading
Alcatel ACC2009: 50% of Reading
Varian PCG 750: +/- 15% of Reading
Varian PVG 500: +/- 15% of Reading