Edwards WRG200 Active Wide Range Gauge | WRG200-x-NW25

The Edwards Active Wide Range Gauge (WRG200) offers the capability of single port pressure measurement in the range atmosphere to 10-9 mbar with a linear output.

Measurement Range: 750.06 to 7.501 × 10⁻¹⁰ Torr
Accuracy: 50 to 1×10-2 mbar ±15% | 1×10-2 to 1×10-8 mbar ±30%

The WRG200 is a compact and efficient solution that significantly reduces the required space and hardware for connectivity. Its compact design makes it ideal for various applications. This gauge has undergone numerous technical improvements such as its innovative striker design, a 3600 LED light ring, easy push-button calibration, integrated set-point controls, and comprehensive diagnostics.

Features and Benefits

  • 360° LED light ring visual pressure indicator
  • Reduced footprint
  • Multipoint star striker
  • Analogue/digital
  • Wide range power supply
  • Drop in compatible
  • New magnets
  • Set-point relay

Applications

The WRG200 offers cost-effectiveness as a vacuum management solution, either when used with an Edwards controller or directly integrated into the system controls. This gauge is perfect for monitoring pressure in systems that are rapidly pumped down from atmospheric pressure to high vacuum levels.

Broad Range Vacuum Gauging

Broad range vacuum gauges are the best choice when the pressure needs to be measured over a wide range and when good accuracy is needed across the whole range. Reliable vacuum gauges are essential to ensure your system’s accuracy and control. Our vacuum gauging solutions include the WRG Series Cold Cathode Combination Gauge, AIM Series Cold Cathode Gauge, and APG Series Pirani Gauge.

DigiVac Recommendations

DASH Portable Capacitance manometer Gauge. The DigiVac Bullseye DASH is the first portable high vacuum display and capacitance manometer gauge in the world; it also powers and displays CM readings and contains patented vacuum graphing technology.

FYRA Vacuum Controller | Multi-Gauge Controller with Touchscreen | ATM to High Vacuum | UL/CSA/CE. The FYRA is a versatile vacuum controller with a touchscreen or dial interface, the ability to measure and control from atmosphere to high vacuum, and the capability to drive different types of sensors. Measurement range is sensor dependent (1000 Torr (10+3) down to 10-10 Torr). Available as Bench top, Panel Mount or Rack Mount.

Downloads

Data Sheet | Brochure

Technical Specifications

General Product Information
Connection Vacuum Inlet Flange DN 25 ISO-KF (NW25)
Weight 0.35 kg
Supply Voltage DC 15-48V
Dimension Length 3.62 in
IP Rating IP40
Supply Frequency DC Voltage
Dimension Height 1.81 in
Dimension Width 1.77 in
dimensionsCombinedLengthWidthHeight 3.62 in
3.62 in
3.62 in
Connection Power RJ45 (FCC 68) – Female
Environmental Conditions
Temperature Ambient Operation Max 50 °C
Temperature Ambient Operation Min. 0 °C
Temperature Ambient Storage Min. -30 °C
Temperature Ambient Storage Max. 70 °C
Classification
Product Pressure Range High Vacuum – 1*10-3 – 1*10-7 mbar
Medium Vacuum – 1 mbar – 1*10-3 mbar
Rough Vacuum 1000 mbar – 1 mbar
Principle Of Measurement Cold-cathode Pirani combination vacuum gauge (IM inverted magnetron)
Material of Construction
Material Body Stainless Steel 1.4404 / 316 L / X2 CrNiMo 17 13 2
Material Exposed To Vacuum Stainless Steel 316L and 304L, Tungsten, Glass, Molybdenum, Trace of Nickel and Nickel iron
Material Filament Tungsten/Rhenium
Operation Conditions – Limitations
Pressure Inlet Continous Max. 1125.09 Torr
Temperature Baking Max. (Elec. Removed) 150 °C
Pressure Inlet Max Overpressure 7500.62 Torr
Temperature Baking Max. 50 °C
Performance Data
Measurement Uncertainty 50 to 1×10-2 mbar ±15%
1×10-2 to 1×10-8 mbar ±30%
Power Consumption Max 4W
Pressure Measurement Max. 750.06 Torr
Pressure Measurement Min.. 7.501 × 10⁻¹⁰ Torr
Variants
Certificated Pressure Measurement None
Connectivity – Communication Interface
Communication Interfaces Customer 0-10 V (analog output – linear)
Connection electrical comms interface connector RJ45 (FCC 68) – Female
Setpoint Quantity 0
Status Indicator LED Light Ring
Communication Interface Protocol 2.0 – 10 V log linear with pressure WRG200

Product Inquiry



WRG200-X-NF25 Active Wide Range Gauge
Edwards WRG200-X-NW25 | Active Wide Range Gauge | 750.06 to 7.501 × 10⁻¹⁰ Torr
$1,581.00