Contact Us:
(732) 765-0900

Description

PKR 360 Cold Cathode Pirani with NW 40

Pfeiffer Compact FullRange® Gauge with Two Sensors PKR 360 Active Pirani Cold Cathode Transmitter KF40

Description

The Pfeiffer PKR 360 is a FullRange® Pirani cold cathode gauge that provides optimal measurements in the range of 1×10-9 to atmosphere.
  • The maximum pressure applies to inert gases and temperatures below 55 °C
  • Rugged and dependable
  • Insensitive to air ingress
  • Without elastomer seal
  • (PN): PT T02 150 010

Full range gauge including a Pirani and cold cathode gauge.

Pirani Features

The PKR 360 features two measurement circuits: Pirani and Cold Cathode. The Pirani circuit is active at all times and the cold cathode circuit becomes active after passing the 1×10-1 Torr threshold, providing a full range of measurement and allowing for use in a wide variety of applications. Vacuum pressure can be displayed via a DCU or Pfeiffer TPG 361 Vacuum Gauge controller.

  • Measurement range:
    • 1 * 10^(-9) to 1000 hPa
  • Accuracy Specifications:
    • 30 % (1E-8 hPa – 1E2 hPa)
  • PKR 360 FullRange® Gauge With Two Sensors in One Housing
  • Optimized for Applications in the Medium Vacuum Range
  • Very Low External Magnetic Field
  • Rugged and Dependable
  • Insensitive to Air Ingress
  • Without Elastomer Seal
  • Features a 40 Millimeter (mm) Diameter (DN 40)
  • Maximum Pressure Refers to Inert Gases and Temperatures of Less Than 55°C

The gauge can not be used for the measurement of highly flammable or combustible gases in the mixture with an oxidizing agent (e.g. atmospheric oxygen) within the explosion limits

 

Specifications

Anode Mo
Connection flange DN 40 ISO-KF
Output signal: Measuring range 2 – 8.6 V
Output signal: Minimum load 10 kΩ
Bakeout temperature max.(electronics removed) ≤ 150 °C
Pressure max. 10000 hPa
Feedthrough Glass
Input voltage(s) 14.5 – 30 V DC
Flange, material Stainless Steel 1.4435 (AISI 316L)
Weight 0.32 kg
Filament Tungsten
Power consumption max. 2 W
Measuring range 1E-9 – 1E3 hPa
Accuracy of measurement in range 1 30 % (1E-8 hPa – 1E2 hPa)
Measurement cable length 300 m
Measuring method
  • Pirani
  • Cold cathode
Temperature: Storage -40 – 70 °C
Ambient temperature 5 – 55 °C
Volume 20.9 cm³
Repeatability in range 1 5 % (1E-8 hPa – 1E2 hPa)

Downloads

Data Sheet 

For more documentation, please visit DigiVac’s folder.

Product Inquiry



pkr360-
PKR 360 | Cold Cathode Pirani | NW 40